GasAlertMax
Pump Calibration
Pump Calibration Pass
If the pump calibration/pump test passes, the audible alarm
beeps three times, the LCD displays rECAL SuccESS, and
the detector returns to normal operation.
Pump Calibration Fail
If the pump calibration/pump test fails, the audible alarm
beeps six times and the LCD displays PumP FAiLurE.
If the pump calibration fails, try to calibrate it again. If the
calibration continues to fail, replace the pump. Refer to
Replacing a Sensor, Pump, or Pump Filter.